Request Information
Download PDF
Configure Output Connector
GTM96600-65VV.V-P2

Information | |||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Model Number | GTM96600-65VV.V-P2 | ||||||||||||||||||||||||||||||||
Description | GTM96600-65VV.V-P2, ITE / Medical Power Supply, 60601-1-4th Ed. , Desktop/External, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60Hz, See "Blade/Cord Installed" section, Output Rating: 65 Watts, 5-54V in 0.1V increments, Approvals: CE China RoHS Double Insulation PSE PSE PSE CB RCM LPS WEEE VCCI Ukraine Level VI EAC CB | ||||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTM96600-TZ/ | ||||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | http://globtek.force.com/models/ec_declaration?id=a0Oa000000PRtgwEAD | ||||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | http://globtek.force.com/models/ec_declaration?id=a0Oa000000PRtgwEAD | ||||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | http://www.globtek.com/pdf/rohs_cert/a0Oa000000PRtgwEAD | ||||||||||||||||||||||||||||||||
REACH Declaration | http://www.globtek.com/pdf/iso_certificates/REACH.pdf | ||||||||||||||||||||||||||||||||
Conflict Minerals Declaration | http://www.globtek.com/pdf/conflict-minerals.pdf | ||||||||||||||||||||||||||||||||
Model Parameters | |||||||||||||||||||||||||||||||||
Type | Desktop/External | ||||||||||||||||||||||||||||||||
Technology | Regulated Switchmode AC-DC Power Supply AC Adaptor | ||||||||||||||||||||||||||||||||
Category | ITE / Medical Power Supply | ||||||||||||||||||||||||||||||||
Input Voltage Range (V) | 100-240V~, 50-60Hz | ||||||||||||||||||||||||||||||||
Max Input Current (A) | 1.5A | ||||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 65 | ||||||||||||||||||||||||||||||||
Output Voltage Range (V) | 5-54 | ||||||||||||||||||||||||||||||||
Efficiency | VI | ||||||||||||||||||||||||||||||||
Ingress Protection | IP41 | ||||||||||||||||||||||||||||||||
Size (mm) | 117.5*53.5*34.2 | ||||||||||||||||||||||||||||||||
Enclosure | |||||||||||||||||||||||||||||||||
Rating Table | |||||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined)
| |||||||||||||||||||||||||||||||||
Specifications | A) ELECTRICAL SPECIFICATIONS: 01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac 90-264 Vac range @ 100% of rated load current 85-264 Vac range @ 85% of rated load current 110-370 VDC range @ 100% of rated load current 02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz 03. Output Regulation: +/- 5% measured at the output connector 04. Line Voltage Regulation: +/- 0.5% typical measured at full load 05. Green Power On Indicator LED 06. Output Ripple (Vp-p): 1% or 100 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector 07. Turn-ON/OFF Overshoot: 5% maximum, 1 mS typical recovery time for 40% to 70% step load 08. Turn-ON Delay: 1 second maximum @ full load, nominal line 09. Hold-Up Time: 8 mS typical @ nominal input voltage and full load 10. Inrush Current: 30A/60A maximum cold start @ 115/230Vac input 11. Efficiency: Compliant with Efficiency Level VI and CoC Tier 2 standards 12. No Load Standby Power: <0.075 W @ 230Vac for <= 49W output power rating, <0.15W @ 230Vac for > 49W output power rating B) PROTECTION 01. Input Protection: Input line fusing and 300Vac MOV 02. Short Circuit/ Overload: Electronically Protected unit will auto recover upon removal of fault Output Current Limit: 110% to 160% of rated output current (upper end of current limit range is determined by normal output power rating, not derated output power level) 03. Output Over-Voltage: 110% to 130% of nominal output voltage C) SAFETY 01. Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output, On Class I models, 3000Vac or 4242Vdc from input to earth 02. Earth Leakage Current: Class I models < 300uA, N/A for Class II models 03. Touch Current: Class I models < 20uA, Class II models < 75uA 04. Earth Continuity Test: < 0.1 Ohm between Earth Pin at AC input and PCB termination point (Class I models only) 05. Means of Protection: 2 x MOPP 06. Compliant Standards: See listings at end of this drawing for specifics D) EMC EN 60601-1-2, 4th edition Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4 Conducted Emissions: Class B, FCC Part 15, Class B Radiated Emissions: Class B, FCC Part 15, Class B Line Frequency Harmonics EN61000-3-2, Class A Voltage Fluctuations/Flicker EN61000-3-3 Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2 Static Discharge Immunity EN61000-4-2, 10kV Contact Discharge, 20kV air discharge Radiated RF Immunity EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM. EFT/Burst Immunity EN61000-4-4, 4kV/100kHz. Line Surge Immunity EN61000-4-5, 2kV differential, 4kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m Voltage Dip Immunity EN61000-4-11, Criteria E) OTHER: 01. MTBF: 1,000,000 Hours @ 40°C ambient temperature, Full Load 02. Operating Temperature: GTM96600 models: -20°C to 40°C ambient temperature at full load, -20°C to 50°C ambient temperature with derating to 80% load 03. Operating Humidity: 0% to 95% relative humidity, non-condensing 04. Storage Temperature: -30°C to 80°C 05. Operating Altitude: 5000 Meters 06. ROHS: Compliant with latest regulations, see approvals section below F) ENCLOSURE 01. Housing: High impact plastic, 94V0 polycarbonate, non-vented 02. Markings: Label or Laser Printed 03. AC Input mechanical options: Desktop C6, C8, C14 or C18 IEC Inlet. Hybrid option (Desktop or Changeable Blade Wall Plug-in) Class I or Class II input G) SPECIAL OPTIONS 01. Cost reduction, removal of LED 02. Custom Cordsets, various cordage types, and connector types 03. Custom Markings 04. Short term Output Surge Capability 05. Reduced Leakage Current versions 06. Tightened output voltage tolerance 07. Reduced output ripple level 08. Reduced output power rating marking 09. High Rel PCB laminate with Plated through Holes for IPC610 Class 2 Compliance 10. Low Temperature operation, down to -40°C (GTM96600) 11. Special Housing Colors and Cordset Colors 12. Quasi Constant Current Output, for Lead-Acid Battery Charger Applications for GTM96300 series H) OUTPUT CONNECTORS Several output connector options are available with various output current ratings. GlobTek can supply 10A rated 2.1mm and 2.5mm style DC Power Jacks, to complement our 10A output rated 2.1mm and 2.5mm DC power plugs used on our output cordsets. 2.1mm PCB DC Power Jack ordering code: JACK-C-PC-10A-RA(R) 2.5mm PCB DC Power Jack ordering code: JACK-L-PC-10A-RA(R) | ||||||||||||||||||||||||||||||||
Derating Curve | |||||||||||||||||||||||||||||||||
Qty (ea.)/Carton | 20 | ||||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | Egg Crate | ||||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | |||||||||||||||||||||||||||||||||
Master Carton Size (mm) | 427 x 405 x 158 | ||||||||||||||||||||||||||||||||
Gross Weight (kg) | 15.6 | ||||||||||||||||||||||||||||||||
Label Width (mm) | 35.0 | ||||||||||||||||||||||||||||||||
Label Height (mm) | 70.0 | ||||||||||||||||||||||||||||||||
Input Configuration | |||||||||||||||||||||||||||||||||
Description | See "Blade/Cord Installed" section | ||||||||||||||||||||||||||||||||
Output Configuration | |||||||||||||||||||||||||||||||||
Common output connector options: Click to add to your request.
For a comprehensive list of options, click here Contact GlobTek for your specific requirements or custom solutions. | |||||||||||||||||||||||||||||||||
Approvals | |||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||
Popular Products | |||||||||||||||||||||||||||||||||
|