GTMG3080P250VV-X.X-FW Open Frame/Internal Medical Power Supply
GTMG3080P250VV-X.X-FW | |||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
GTMG3080P250VV-X.X-FW, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 250 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: CE; China RoHS; cUR; GOST-R; PSE; VCCI; RoHS; WEEE; Double Insulation; Ukraine; cULus; | |||||||||||||||||||||||||||||||
Information | |||||||||||||||||||||||||||||||
Model Number | GTMG3080P250VV-X.X-FW | ||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTMG3080PWWWVV-X.X-FW/ | ||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | Contact GlobTek for more information | ||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ec_declaration?id=a0Oa000000MEd16EAD | ||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||
UKCA-Declaration for Medical | Contact GlobTek for more information | ||||||||||||||||||||||||||||||
UKCA-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ukca_declaration?id=a0Oa000000MEd16EAD | ||||||||||||||||||||||||||||||
UKCA-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||
Morocco SDoC Declaration | https://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MEd16EAD | ||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | https://www.globtek.com/pdf/rohs_cert/a0Oa000000MEd16EAD | ||||||||||||||||||||||||||||||
Prop 65 Statement | https://www.globtek.com/r2/PH7CNf | ||||||||||||||||||||||||||||||
REACH Declaration | https://www.globtek.com/r2/eM2co8 | ||||||||||||||||||||||||||||||
Conflict Minerals Declaration | https://www.globtek.com/r2/vZY8R5 | ||||||||||||||||||||||||||||||
Model Parameters | |||||||||||||||||||||||||||||||
Type | Open Frame/Internal | ||||||||||||||||||||||||||||||
Technology | Open Frame Medical Power Supply | ||||||||||||||||||||||||||||||
Category | Medical Power Supply | ||||||||||||||||||||||||||||||
Input Voltage Range (V) | 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz) | ||||||||||||||||||||||||||||||
Max Input Current (A) | 3.23A / 100VAC, 1.23A / 240VAC | ||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 250 | ||||||||||||||||||||||||||||||
Output Voltage Range (V) | 12-48 | ||||||||||||||||||||||||||||||
Efficiency | N/A | ||||||||||||||||||||||||||||||
Ingress Protection | IPX0 | ||||||||||||||||||||||||||||||
Size (mm) | 76.2x39.6x127 | ||||||||||||||||||||||||||||||
Enclosure | |||||||||||||||||||||||||||||||
Rating Table | |||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined) | |||||||||||||||||||||||||||||||
Specifications | |||||||||||||||||||||||||||||||
250W Open Frame Medical Power Supply Features: Universal AC input / Full range with active PFC 5"x 3" footprint Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1; IEC 60601-1 / 2xMOPP 150W convection cooling and 250W with 23.5 CFM forced air Protection: Short circuit, Over load, Over voltage High operation temperature up to 70°C Power OK signal Built-in fan supply Withstand 2G vibration test ELECTRICAL SPECIFICATIONS: Output DC Voltage Range: 12V 15V 24V 48V Rated Curren:t 20.85A 16.67A 10.45A 5.25A Current Range (Convection): 0 ~ 12.5A 0 ~ 10A 0 ~ 6.25A 0 ~ 3.125A Current Range (23.5 CFM Fan): 0 ~ 20.85A 0 ~ 16.67A 0 ~ 10.45A 0 ~ 5.25A Rated Power (Convection): 150W 150W 150W 150W Rated Power (23.5 CFM Fan): 250.2W 250W 250.8W 252W Ripple & Noise (Max.): 120mVp-p 150mVp-p 240mVp-p 480mVp-p Voltage Adj. Range: ±10% Voltage Tolerance: ±2.0% ±2.0% ±1.0% ±1.0% Line Regulation: ±0.5% ±0.5% ±0.5% ±0.5% Load Regulation: ±2.0% ±2.0% ±2.0% ±2.0% Setup, Rise Time: 2000ms, 50ms / 230VAC, 3500ms, 50ms / 115VAC at full load Hold Up Time: >10ms / 230VAC, >10ms / 115VAC at full load Input Voltage Range: 90 ~ 264VAC, 127 ~ 373VDC Frequency Range: 47 ~ 63Hz Efficiency (Typ.) at 230VAC: 90% 90% 91% 91% AC Current (Typ.): 3.23A / 100VAC, 1.23A / 240VAC Inrush Current (Typ.): Cold Start 30A / 115VAC, 60A / 230VAC Leakage Current: For earth < 300uA / 264VAC, For patient < 100uA / 264VAC Protection Over Load: Above 110% rated output power Protection type: latch-style (Recovery after reset AC power ON) Over Voltage: 115 ~ 140% rated output voltage Protection type: latch-style (Recovery after reset AC power ON) Environment Working Temp.: -20 ~ +70°C (Refer to output load de-rating curve) Working Humidity: 20 ~ 90% RH non-condensing Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH Temp. Coefficient: ±0.03% / °C (0 ~ 50°C) Vibration: 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes Safety & EMC Safety Standards: Meets UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99 Meets TUV (EN 60601-1; IEC 60601-1); MOPP approval Withstand Voltage: I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC Isolation Resistance: I/P-O/P: 100M Ohms / 500VDC EMI Conduction & Radiation: Certified EN 55011 Harmonic Current: Certified EN 61000-3-2; EN 61000-3-3 EMS Immunity: Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11 Others MTBF: XXX K HRS Certified MIL-HDBK-217F Cooling: Free Air convection for 150W, with 23.5 CFM Fan for 250W Dimension (WxHxH): 76.2x39.6x127 mm / 3.00x1.43x5.00 inch Packing: 0.32 kg Note 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor. 3. Tolerance: includes set up tolerance, line regulation and load regulation. 4. Line regulation is measured from low line to high line at rated load. 5. Load regulation is measured from 0% to 100% rated load. 6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. | |||||||||||||||||||||||||||||||
Derating Curve | |||||||||||||||||||||||||||||||
Qty (ea.)/Carton | TBD | ||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | 130*77*45 | ||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | White | ||||||||||||||||||||||||||||||
Master Carton Size (mm) | TBD | ||||||||||||||||||||||||||||||
Gross Weight (kg) | TBD | ||||||||||||||||||||||||||||||
Label Width (mm) | TBD | ||||||||||||||||||||||||||||||
Label Height (mm) | TBD | ||||||||||||||||||||||||||||||
Input Configuration | |||||||||||||||||||||||||||||||
Description | |||||||||||||||||||||||||||||||
Approvals | |||||||||||||||||||||||||||||||
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice. Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation. | |||||||||||||||||||||||||||||||
|