Request Information Download PDF
GTM41134-WWVV-FWT2 GTM41134-WWVV-FWT2
Information
Model NumberGTM41134-WWVV-FWT2
DescriptionGTM41134-WWVV-FWT2, Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C8 AC Inlet connector, Class II, Non-Earth Ground (aka "Figure-8"), Output Rating: 6 Watts, 3.3-48V in 0.1V increments, Approvals: LPS CE
Agency Documentshttp://www.globtek.info/certs/GTM41134/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000MEmfkEAD
CE EC-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
RoHS/RoHS2 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000MEmfkEAD
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC
CategoryMedical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)0.6A Max
Wattage (W)6
Vout Range (V)3.3-48
EfficiencyN/A
Ingress Protection
Size (mm)84.5*41.5*17.0
EnclosureUser-added image
 
Rating Table
Model NumberVoltageAmps (A)Watts (W)RFQ
GTM41134-0606-1.0-FWT251.26RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 1% typical measured at full load
05. Output Ripple (Vp-p): +/- 1% of output voltage, Vp-p, typical
06. Turn-ON/OFF: 5% maximum,1 mS typical recovery time for 25% step load
07. Turn-ON Delay: 1 Sec, typical
08. Hold-Up Time: 8 mS typical @ nominal input voltage and full load
09. Inrush Current: 30A maximum cold start @ 115 Vac input, 60A maximum cold start @ 230 Vac input
10. Switching Frequency: 100 KHz typical, varies with line and load
11. Efficiency: 75% typical at nominal input and full load

B) PROTECTION
01. Over-Voltage: limited via crowbar method
02. Short Circuit/Over-load: momentary short circuit protected, output will autorecover upon removal of short
03. Input Protection: Input line

C) SAFETY
01. Dielectric Withstand Voltage: 5656 Vdc from primary to secondary
02. Earth Leakage Current: Did not exceed 5 mA in NORMAL CONDITION and 10 mA in SINGLE FAULT CONDITION
03. Touch Current: Did not exceed 100uA NC(Normal condition) 500uA SFC(single fault condition)
04. Earth continuity test:<0.1 Ohm or < 0.2 Ohm when with a non-detachable Power Supply cord, Class I Only
05. Type of protection against electric shock: class i
06. Degree of safety of application in the presence of flammable anesthetic: not evaluated
07. Mode of operation: Continuous
08. ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006


D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature: 0°C to 40°C ambient temperature
03. Humidity: 0% to 90% relative humidity
04. Storage Temperature: -10°C to 80°C
05. Cooling: Convection

E) ENCLOSURE
01. Housing: Open Frame
02. Size: 84.6 x 41.4 x 28.9 mm
Derating Curve
Qty (ea.)/CartonTBD
Individual Box/Gift Box Size (mm)TBD
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)TBD
Label Width (mm)25.4
Label Height (mm)17.8
Input Configuration
DescriptionIEC 60320/C8 AC Inlet connector, Class II, Non-Earth Ground (aka "Figure-8")User-added image

Mates with IEC 60320/C7 Plug

Standard International IEC 320/C7 Cordsets

Below are standard cordsets which are "not included" (unless stated above); these may be purchased separately or packaged with the power supply. Please contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.
 

2 Conductors (No Ground) IEC 320/C7 (Figure "8")

4511116F703A(R)

N. American

NEMA 1-15P

IEC 320/C7

1830

6

5014112M703A(R)

Australian

AS 3112

IEC 320/C7

2000

7

4533501M8703(R)

China

GB 2099.1

IEC 320/C7

1830

6

2074112M703A(R)

European

CEE 7/16

IEC 320/C7

2000

7

451J3401M8703(R)

Japan

JIS 8303

IEC 320/C7

1830

6

6104112M703A(R)

UK

BS1363

IEC 320/C7

2000

7

Approvals
LogoDescription
User-added image  CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010