GTM91110P240VV-CAW-S Enclosed-Metal ICT / ITE / Medical Power Supply

GTM91110P240VV-CAW-S
GTM91110P240VV-CAW-S, ICT / ITE / Medical Power Supply, Enclosed-Metal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Molex 26-60-4030  3 Position Header Pin 1: Line, Pin 2: Removed, PIN 3: Neutral, Output Rating: 200 Watts, Power rating with convection cooling (W) , 12-55V in 0.1V increments, Approvals: Book 60601; CB 60601-1; CE; cETLus 60601-1 3rd; WEEE; VCCI; CE; China RoHS; Double Insulation; FCC; GOST-R; RoHS; SIQ; SIQ; Ukraine;GTM91110P240VV-CAW-S, ICT / ITE / Medical Power Supply, Enclosed-Metal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Molex 26-60-4030 3 Position Header Pin 1: Line, Pin 2: Removed, PIN 3: Neutral, Output Rating: 200 Watts, Power rating with convection cooling (W) , 12-55V in 0.1V increments, Approvals: Book 60601; CB 60601-1; CE; cETLus 60601-1 3rd; WEEE; VCCI; CE; China RoHS; Double Insulation; FCC; GOST-R; RoHS; SIQ; SIQ; Ukraine;
Information
Model NumberGTM91110P240VV-CAW-S
Agency Documentshttp://www.globtek.info/certs/GTM91110P/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000MgYsMEAV
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000MgYsMEAV
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MgYsMEAV
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000MgYsMEAV
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeEnclosed-Metal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)4.0A
Max Output Power Rating (W)200
Output Voltage Range (V)12-55
EfficiencyN/A
Ingress Protection
Size (mm)140*87.6*51.9
Enclosure   
User-added image



                                            









                                                       Label 1
                                  Position Label 1 on outside bottom plate


                                                       Label 2
Material: White art paper with Black nomenclature and is attached on top of CX1

                          User-added image

                                                     Label 3
Material: White art paper with Black nomenclature and is attached on side of CX1
                         User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM91110P24048-CAW-S*484.17200.16
GTM91110P24036-6.0-CAW-S*306.67200.1
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50/60 Hz
03. Inrush current: Cold start @ 25°C, 40A @ 115Vac, 80A @ 230Vac
04. Efficiency: 85% to 90% typical at full load, 120-240Vac input
05. Power factor: 0.98 typical at full load, 230Vac input
06. Turn on delay: 5 Seconds maximum, (3 seconds max option)
07. Load regulation: ± 1% Measured at the connector side
08. Line regulation: ± 0.2% Measured at the connector side
09. Output Ripple protection (Peak to Peak): +/-1% maximum of nominal output voltage measured at 20 MHz bandwidth with 0.1µf ceramic capacitor in parallel with a low impedance 47 µf electrolytic capacitor located 200mm from the end  of the output connector
10. Turn on/off overshoot: 5% Maximum, 1mS recovery time
11. Hold up time: 8 mS @ 300W load, 10mS@200W load, 16mS@150W load at 115Vac input
12. Transient response: 2% Maximum deviation, returns to initial condition in 3 ms max for 25% load step,   from 10% to 100% of rated load
13. Remote sense (optional): compensates for up to 0.5V voltage drop in output wiring.
Power supply will operate correctly, with or without remote sense pins hooked up.
14. Auxillary Output: For forced air fan 12V @ 250mA, if convection cooled 12V @ 150mA. Other output voltages and current ratings from 5V to 24V available, contact factory for availability.
15. Indicator: Green LED for output presence

B) PROTECTION
01. Input Protection: Input line fusing and 300Vac MOV
02. Short circuit: Protected. Unit will autorecover upon removal of short
03. Overload protection:  Autorecover upon removal of overload
    150W, 175% - 210% of nominal power rating
    200W, 130% - 160% of nominal power rating
    240W, 110% - 130% of nominal power rating
    300W, 110% - 130% of nominal power rating
04. Over voltage: Protected at 110- 140% of nominal output voltage, latched type. AC cycling required to restore output. Autorecovery type OVP is an option, contact factory for availability.
05. Fan output protection: 3 Pin Linear Regulator, short circuit proof

C) SAFETY
01. Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output,
     On Class I models, 2500Vac or 3535Vdc from input to earth
02. Earth Leakage Current:
     Class I Input Models: Does not exceed 500uA
     Class II Input Models: Does not apply
03. Touch Current: Does not exceed 100uA, output terminations to earth
04. Output terminations rated as SELV, and Means of Operator Protection

D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature Range:
        -10°C to 50°C @ 100% output power rating
        -10°C to 70°C @ 50% output power derating
03. Cooling:
          Open Frame 150W Series: Convection
          Open Frame 240W and 300W Series: 20 CFM Forced Air Cooling
          Perforated Plastic Housing 200W Series: 20 CFM Forced Air Cooling,
                   with convection cooling 50% derated (100W) output capability.
04. Relative Humidity: operating 0% to 90% Non-condensing
05. Storage Temperature: -30°C to 85°C
06. Altitude: up to 2000 meter
07. Random vibration/transient drop: MIL-STD 810E

E) MECHANICAL  FORMAT:
150W, 240W and 300W Open Frame version
      Size: 127*76.2*43 mm
      Weight: 470 grams
200W Perforated Plastic Enclosure version
      Size: 140*87.6*51.9 mm
      Weight: 620 grams

 
Derating CurveUser-added image
Qty (ea.)/Carton36
Individual Box/Gift Box Size (mm)160*67*90
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)430 x 340 x 304
Gross Weight (kg)14.0
Label Width (mm)50.0
Label Height (mm)50.0
Input Configuration
DescriptionMolex 26-60-4030 3 Position Header Pin 1: Line, Pin 2: Removed, PIN 3: Neutral
User-added image     
                                                     User-added image
    
Optional Input Wiring Kits are available:
PN/Ordering No   Length  End A   End B
XHP410070M7MLX3(R)70JST XHP-4  Molex 09-93-0300
C810070M6MLX3(R)60IEC60320 C8  Molex 09-93-0300
C1410070M6MLX3(R)60IEC60320 C14  Molex 09-93-0300
 Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
No Logo ApplicableCB report IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) and or EN 60601-1:2006 3rd Edition including 2MOOP requirements
User-added image
 
cETLus-Cert-4007497 ANSI/AAMI ES60601-1:2005/A1:2012 Issued: 2012/08/20 Medical electrical equipment ? Part 1: Amendment 1; CAN/CSAC22.2 No. 60601-1:14, Third Edition Issued: 2014/03/01 - Medical Electrical Equipment - Part 1
User-added image  CE Mark: tested to comply with EN55022:2006/A1:2007 Class B, EN610003-2, EN610003-3 including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6N EN61000-4-11; EMI: Complies with EN55011 CLASS B and FCC Part 15B - On label or Molded in case
User-added image60601-1 3rd Operating Instructions
User-added imagenull
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
User-added image
FCC PART 15 CLASS B
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added imageNote for Maximum Power GTM91110P240-CA(W)-S
User-added image  CE Mark: tested to comply with EN61000-3-2, EN61000-3-3 and EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61001-1-2-2007 EMI: Complies with EN55011 CLASS B and FCC Part 15 B
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added image
IEC/EN60950-1
SIQ Mark License EN60950-1
User-added image
IEC/EN60601-1
SIQ Mark License EN60601-1
Request Info