GTMG3074P-05VV-X.X-FW Open Frame/Internal ICT / ITE / Medical Power Supply

GTMG3074P-05VV-X.X-FW
GTMG3074P-05VV-X.X-FW, ICT / ITE / Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 85 ~ 264VAC, 125 ~ 373VDC (47 ~  63Hz), Solder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins, Output Rating: 5 Watts, Power rating with convection cooling (W) , 5 ~ 24V in 0.1V increments, Approvals: cUR; PSE; CE; China RoHS; WEEE; GOST-R; cULus; VCCI; Double Insulation; Ukraine; RoHS;GTMG3074P-05VV-X.X-FW, ICT / ITE / Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 85 ~ 264VAC, 125 ~ 373VDC (47 ~ 63Hz), Solder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins, Output Rating: 5 Watts, Power rating with convection cooling (W) , 5 ~ 24V in 0.1V increments, Approvals: cUR; PSE; CE; China RoHS; WEEE; GOST-R; cULus; VCCI; Double Insulation; Ukraine; RoHS;
Information
Model NumberGTMG3074P-05VV-X.X-FW
Agency Documentshttp://www.globtek.info/certs/GTMG3074-WVV-X.X-FW/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000MEi3CEAT
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000MEi3CEAT
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MEi3CEAT
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000MEi3CEAT
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyOpen Frame Medical Power Supply
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)85 ~ 264VAC, 125 ~ 373VDC (47 ~ 63Hz)
Max Input Current (A)0.13A / 115VAC, 0.08A / 230VAC
Max Output Power Rating (W)5
Output Voltage Range (V)5 ~ 24
EfficiencyN/A
Ingress ProtectionIPX0
Size (mm)45x19.1x57.9
EnclosureUser-added image
 
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTMG3074P-0505-FW*515
GTMG3074P-0512-FW*120.425.04
GTMG3074P-0515-FW*150.334.95
GTMG3074P-0524-FW*240.215.04
* (Final ordering #/Part No to be defined)
Specifications
Features:
Universal AC input / Full range
Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1: 2006; IEC 60601-1 / 2xMOPP
Isolation class Ⅱ
Green design, No-load power consumption < 0.3W Protection: Short circuit, Over load, Over voltage
Brown-out (Low AC Input Voltage) High operation temperature up to 70°C Withstand 2G vibration test
3 years warranty


ELECTRICAL SPECIFICATIONS:
Output
DC Voltage Range: 5V 12V 15V 24V
Rated Current: 1A 0.42A 0.33A 0.21A
Current Range: 0 ~ 1A 0 ~ 0.42A 0 ~ 0.33A 0 ~ 0.21A
Rated Powerv 5W 5.04W 4.95W 5.04W
Ripple & Noise (Max.)v 100mVp-p 150mVp-p 150mVp-p 240mVp-p
Voltage Adj. Rangev ±10%
Voltage Tolerancev ±2% ±1% ±1% ±1%
Line Regulation: ±0.5% ±0.5% ±0.5% ±0.5%
Load Regulationv ±1% ±0.5% ±0.5% ±0.5%

Input
Voltage Range: 85 ~ 264VAC, 125 ~ 373VDC
Frequency Range: 47 ~ 63Hz
Efficiency (Typ.) at 230VAC: 71% 75% 75% 77%
AC Current (Typ.): v
Inrush Current (Typ.): Cold Start 25A / 115VAC, 45A / 230VAC

Protection
Over Loadv Above 110% rated output power, Protection type: Hiccup mode, recovers automatically after fault condition is removed
Over Voltagev 115 ~ 145% rated output voltage, Protection type: latch-off mode


Environment
Working Temp.: -20 ~ +70°C (Refer to load de-rating curve)
Working Humidity: 20 ~ 90% RH non-condensing
Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH
Temp. Coefficient: ±0.03% / °C (0 ~ 50°C)
Vibration: 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes

Safety & EMC
Safety Standards: Certified UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99 Certified TUV (EN 60601-1; IEC 60601-1); MOPP approval
Withstand Voltage: I/P-O/P: 4KVAC
Isolation Resistance: I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation: Certified EN 55011
Harmonic Current: Certified EN 61000-3-2; EN 61000-3-3
EMS Immunity: Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11

Others
MTBF : 217.4K HRS Certified MIL-HDBK-217F
Dimension (WxHxD): 45x19.1x57.9 mm / 1.771x0.752x2.2798 inch
Packing: 0.033kg; 144pcs / 5.75kg / 1.43CUFT

1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes setup time tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
7. Length of setup time is measured at first cold start. Turning ON/OFF the power supply continuously may increase the setup time.
Derating CurveUser-added image

 
Qty (ea.)/Carton144
Individual Box/Gift Box Size (mm)59*20*47
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)5.75
Label Width (mm)30.0
Label Height (mm)20.0
Input Configuration
DescriptionSolder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
User-added image  CE Mark: tested to comply with EN61000-3-2, EN61000-3-3 and EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61001-1-2-2007 EMI: Complies with EN55011 CLASS B and FCC Part 15 B
User-added imageJAPAN TUV Rheinland-PSE GlobTek Inc to J60950-1(H26) , J55022(H22),J3000(H25).Please follow the procedure listed in the following link for proper import to Japan: https://www.globtek.com/r2/Szj4Vb
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
User-added imageUR 60601-1, 3rd Edition, CAN/CSA-C22.2 No. 601.1-1
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageUKCA Certification
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imagenull
User-added imageUL60950-1, cULus to UL 60950-1 E170507
Request Info