GTMG3069-05VV-X.X-PP Encapsulated in plastic housing Medical Power Supply

GTMG3069-05VV-X.X-PP
GTMG3069-05VV-X.X-PP, Medical Power Supply, Encapsulated in plastic housing, Modular Medical Power Supply, , Input Rating: 85 ~ 264VAC, 125 ~ 373VDC (47 ~  63Hz), Solder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins, Output Rating: 5 Watts, Power rating with convection cooling (W) , 5-24V in 0.1V increments, Approvals: CE; China RoHS; WEEE; GOST-R; Ukraine; VCCI; RoHS; cUR; PSE; Double Insulation; cULus;GTMG3069-05VV-X.X-PP, Medical Power Supply, Encapsulated in plastic housing, Modular Medical Power Supply, , Input Rating: 85 ~ 264VAC, 125 ~ 373VDC (47 ~ 63Hz), Solder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins, Output Rating: 5 Watts, Power rating with convection cooling (W) , 5-24V in 0.1V increments, Approvals: CE; China RoHS; WEEE; GOST-R; Ukraine; VCCI; RoHS; cUR; PSE; Double Insulation; cULus;
Information
Model NumberGTMG3069-05VV-X.X-PP
Agency Documentshttp://www.globtek.info/certs/GTMG3069-WWVV-X.X-PP/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000MEhAcEAL
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000MEhAcEAL
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MEhAcEAL
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000MEhAcEAL
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeEncapsulated in plastic housing
TechnologyModular Medical Power Supply
CategoryMedical Power Supply
Input Voltage Range (V)85 ~ 264VAC, 125 ~ 373VDC (47 ~ 63Hz)
Max Input Current (A)0.13 A / 115VAC, 0.08A / 230VAC
Max Output Power Rating (W)5
Output Voltage Range (V)5-24
EfficiencyN/A
Ingress ProtectionIPX0
Size (mm)50.0x19.7x62.85
EnclosureUser-added image
 
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTMG3069-0505-PP*515
GTMG3069-0512-PP*120.425.04
GTMG3069-0515-PP*150.334.95
GTMG3069-0524-PP*240.215.04
* (Final ordering #/Part No to be defined)
Specifications
Features:
Universal AC input / Full range
Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1: 2006; IEC 60601-1 / 2xMOPP
Isolation class Ⅱ
Green design, No-load power consumption < 0.3W Protection: Short circuit, Over load, Over voltage
Brown-out (Low AC Input Voltage) High operation temperature up to 70°C Withstand 2G vibration test
3 years warranty

ELECTRICAL SPECIFICATIONS:

output
DC Voltage Range: 5V 12V 15V 24V
Rated Current: 1A 0.42A 0.33A 0.21A
Current Range: 0 ~ 1A 0 ~ 0.42A 0 ~ 0.33A 0 ~ 0.21A
Rated Power: 5W 5.04W 4.95W 5.04W
Ripple & Noise (Max.): 2 100mVp-p 150mVp-p 150mVp-p 240mVp-p
Voltage Tolerance: +5 / -3%
Line Regulation: Note.8 ±0.5%
Load Regulation: Note.8 ±1% ±0.5% ±0.5% ±0.5%
Setup, Rise Time: 100ms, 25ms / 230VAC, 100ms, 25ms / 115VAC at full load
Hold Up Time: 100ms / 230VAC, 25ms / 115VAC at full load

Input
Voltage Range: 85 ~ 264VAC, 125 ~ 373VDC
Frequency Range: 47 ~ 63Hz
Efficiency (Typ.) at 230VAC: 71% 75% 75% 77%
AC Current (Typ.): 0.13 A / 115VAC, 0.08A / 230VAC
Inrush Current (Typ.): Cold Start 25A / 115VAC, 45A / 230VAC

Protection
Over Load: Above 110% rated output power Protection type: Hiccup mode, recovers automatically after fault condition is removed
Over Voltage: 115 ~ 145% rated output voltage, Protection type: latch-off mode

Environment
Working Temp.: -20 ~ +70°C (Refer to output load de-rating curve)
Working Humidity: 20 ~ 90% RH non-condensing
Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH
Temp. Coefficient: ±0.03% / °C (0 ~ 50°C)
Vibration 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes

Safety & EMC
Safety Standards: Certified UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99
Certified TUV (EN 60601-1; IEC 60601-1); MOPP approval

Withstand Voltage: I/P-O/P: 4KVAC
Isolation Resistance:I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation: Certified EN 55011
Harmonic Current: Certified EN 61000-3-2; EN 61000-3-3
EMS Immunity: Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11

Others
MTBF: 217.4K HRS Certified MIL-HDBK-217F
Dimension (WxHxH): 50.0x19.7x62.85 mm / 1.969x0.776x2.474 inch
Packing: 0.08kg; 144pcs / 12.52kg / 1.43CUFT


1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes setup time tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 10% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
7. Length of setup time is measured at first cold start. Turning ON/OFF the power supply continuously may increase the setup time.
8.The line regulation and load regulation are calculated based on output voltage.
Derating CurveUser-added image

 
Qty (ea.)/Carton144
Individual Box/Gift Box Size (mm)70*55*27
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)12.52
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionSolder Pins 2 * Ø1.05 mm, 3.5 mm Long, 20 mm Center to Center between Pins
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
User-added imageJAPAN TUV Rheinland-PSE GlobTek Inc to J60950-1(H26) , J55022(H22),J3000(H25).Please follow the procedure listed in the following link for proper import to Japan: https://www.globtek.com/r2/Szj4Vb
User-added imageUL60950-1, cULus to UL 60950-1 E170507
User-added imageUR 60601-1, 3rd Edition, CAN/CSA-C22.2 No. 601.1-1
User-added imagenull
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added imageIndoor Use Only - Mark is on the label or Molded in the case

User-added image  
CE Certification
Request Info