GTMG3079P450VV-X.X-F Open Frame/Internal Medical Power Supply

GTMG3079P450VV-X.X-F
GTMG3079P450VV-X.X-F, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector
Mating Housing: JST VHR-3N or equivalent
Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 450 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: CE; China RoHS; WEEE; Class I; RoHS; VCCI; Ukraine; GOST-R; PSE; cUR; cULus;GTMG3079P450VV-X.X-F, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 450 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: CE; China RoHS; WEEE; Class I; RoHS; VCCI; Ukraine; GOST-R; PSE; cUR; cULus;
Information
Model NumberGTMG3079P450VV-X.X-F
Agency Documentshttp://www.globtek.info/certs/GTMG3079PWWWVV-X.X-F/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000MEctoEAD
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000MEctoEAD
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MEctoEAD
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000MEctoEAD
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyOpen Frame Medical Power Supply
CategoryMedical Power Supply
Input Voltage Range (V)90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz)
Max Input Current (A)4.07A / 100VAC, 1.64A / 240VAC
Max Output Power Rating (W)450
Output Voltage Range (V)12-48
EfficiencyN/A
Ingress ProtectionIPX0
Size (mm)101.6x39.6x177.8
Enclosure
User-added imageUser-added imageUser-added image
User-added image
 
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTMG3079P45012-F*1237.5450
GTMG3079P45015-F*1530450
GTMG3079P45024-F*2418.75450
GTMG3079P45048-F*489.38450.24
* (Final ordering #/Part No to be defined)
Specifications
Features:
Universal AC input / Full range with active PFC
5"x 3" footprint
Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1; IEC 60601-1 / 2xMOPP
300W convection cooling and 450W with 23.5 CFM forced air
Protection: Short circuit, Over load, Over voltage
High operation temperature up to 70°C Built-in fan auxiliary Power
Power OK signal
Remote ON / OFF, Remote sense function
Withstand 2G vibration test




ELECTRICAL SPECIFICATIONS:

Output
DC Voltage Range: 12V 15V 24V 48V
Rated Current: 37.5A 30A 18.75A 3.98A
Current Range (Convection): 0 ~ 25A 0 ~ 20A 0 ~ 12.5A 0 ~ 6.25A
Current Range (23.5 CFM Fan)v 0 ~ 37.5A 0 ~ 30A 0 ~ 18.75A 0 ~ 9.38A
Rated Power (Convection)v 300W 300W 300W 300W
Rated Power (23.5 CFM Fan)v 450W 450W 450W 450.2W
Ripple & Noise (Max.)v 150mVp-p 150mVp-p 240mVp-p 480mVp-p
Voltage Adj.v Range ±5%
Voltage Tolerance: ±2.0% ±2.0% ±2.0% ±2.0%
Line Regulation: ±1.0% ±1.0% ±1.0% ±1.0%
Load Regulation: ±1.0% ±1.0% ±1.0% ±1.0%
Setup, Rise Time: 800ms, 150ms / 230VAC at full load
Hold Up Time: >16ms at 450W / 230VAC

Input
Voltage Range: 90 ~ 264VAC, 127 ~ 373VDC
Frequency Range: 47 ~ 63Hz
Efficiency (Typ.) at 230VAC: 90% 90% 90% 91%
AC Current (Typ.)v 5.23A / 100VAC, 2.09A / 240VAC
Inrush Current (Typ.): Cold Start 40A / 115VAC, 80A / 230VAC
Leakage Currentv For earth < 300uA / 264VAC, For patient < 200uA / 264VAC
Protection
Over Current / Short Circuit: Constant current limit: > 105% rated power, Auto-recovery: If O/P drop to 40% of the rated output voltage, PSU will shut down and auto-recover
Over Voltage: 115 ~ 140% rated output voltage, Protection type: latch-off mode

Environment
Working Temp.v -20 ~ +70°C (Refer to output load de-rating curve)
Working Humidity: 20 ~ 90% RH non-condensing
Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH
Temp. Coefficient: ±0.03% / °C (0 ~ 50°C)
Vibration: 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes

Safety & EMC
Safety Standards: Meets UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99, Meets TUV (EN 60601-1; IEC 60601-1); MOPP approval
Withstand Voltage: I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC
Isolation Resistance: I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation: Certified EN 55011
Harmonic Current: Certified EN 61000-3-2; EN 61000-3-3
EMS Immunity: Certified Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11

Others
Cooling: Free Air convection for 300W, with 23.5 CFM Fan for 450W
Dimension (WxHxH): 101.6x39.6x177.8 mm / 4.00x1.56x7.00 inch
Packing: 0.7 kg
Derating CurveUser-added image
 
Qty (ea.)/CartonTBD
Individual Box/Gift Box Size (mm)180*40*103
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)TBD
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
Description3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
nullClass I Earth Ground
User-added imageUL60950-1, cULus to UL 60950-1 E170507
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf

User-added image  
CE Certification
User-added imageUR 60601-1, 3rd Edition, CAN/CSA-C22.2 No. 601.1-1
User-added imageUKCA Certification
User-added imageJAPAN TUV Rheinland-PSE GlobTek Inc to J60950-1(H26) , J55022(H22),J3000(H25).Please follow the procedure listed in the following link for proper import to Japan: https://www.globtek.com/r2/Szj4Vb
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
Request Info